Thank you for visiting nature.com. You are using a browser version with limited support for CSS. To obtain the best experience, we recommend you use a more up to date browser (or turn off compatibility mode in Internet Explorer). In the meantime, to ensure continued support, we are displaying the site without styles and JavaScript.
Scientific Reports volume 13, Article number: 18407 (2023 ) Cite this article High Accuracy Measurement
Mediastinal structure measurements are important for the radiologist’s review of computed tomography pulmonary angiography (CTPA) examinations. In the reporting process, radiologists make measurements of diameters, volumes, and organ densities for image quality assessment and risk stratification. However, manual measurement of these features is time consuming. Here, we sought to develop a time-saving automated algorithm that can accurately detect, segment and measure mediastinal structures in routine clinical CTPA examinations. In this study, 700 CTPA examinations collected and annotated. Of these, a training set of 180 examinations were used to develop a fully automated deterministic algorithm. On the test set of 520 examinations, two radiologists validated the detection and segmentation performance quantitatively, and ground truth was annotated to validate the measurement performance. External validation was performed in 47 CTPAs from two independent datasets. The system had 86–100% detection and segmentation accuracy in the different tasks. The automatic measurements correlated well to those of the radiologist (Pearson’s r 0.68–0.99). Taken together, the fully automated algorithm accurately detected, segmented, and measured mediastinal structures in routine CTPA examinations having an adequate representation of common artifacts and medical conditions.
Sergei P. Primakov, Abdullah Ibrahim, ... Philip Lambin
Aydin Demircioğlu, Magdalena Charis Stein, … Kai Nassenstein
Dorothea Theil, David Petschelt, ... Dominik Geisel
Computed tomography (CT) pulmonary angiography (CTPA) is a medical imaging procedure where an intravenous (IV) contrast agent is used to enhance visualization of the thoracic blood vessels, in particular the pulmonary arteries1. The CTPA is the current reference imaging method for diagnosis of pulmonary embolism (PE)2,3, a condition associated with high mortality and morbidity that in the US alone affects more than 600,000 patients per year and causes > 100,000 deaths annually4,5. In the PE CTPA reading and reporting process radiologists manually perform quality control and different measurements of mediastinal structures in the imagery depending on the clinical situation and local tradition. Measurements of standard deviation (SD) of Hounsfield Units (HU) in a given region of interest (ROI) such as the descending aorta (DAo) and mean value of HU in the pulmonary trunk (PT), can be used for image quality assessment6,7. The mean value of HU in PT can help choosing an optimal window setting for PE detection8. Ascending aorta (AAo) and PT diameters could alert the radiologist to an aneurysm9,10 or possible pulmonary hypertension11,12,13,14. Performing these manual measurements is time-consuming15, and fully automatic solutions could therefore be of value to the radiologist.
Automated measurements of mediastinal structures demand accurate algorithms for detection and segmentation. Several solutions have been proposed to semi-automatically segment and measure mediastinal structures in CTPA examinations16,17,18. Deterministic approaches such as model-based18 and iterative19 methods for segmenting large vessels in CTPA examinations have been demonstrated. More recently, probabilistic approaches such as deep learning based systems have emerged17,20. However, these studies suffer from two main limitations in their datasets; they either (i) contain only a small number of examinations or (ii) lack examinations having the artifacts and co-morbidities encountered in daily radiology practice. Deep learning-based approaches have been trained and tested with a large number of CTPA examinations, but there is insufficient information on dataset characteristics21. To overcome these limitations, there is a need to train and test algorithms on large datasets with examinations of different image quality containing representative artifacts and medical conditions often encountered in radiology practice. Hence, in this study, we aimed to develop an algorithm that can accurately detect, segment and measure mediastinal structures in routine clinical CTPA examinations and benchmark its performance to the radiologist.
A total of 700 CTPA examinations performed on 652 patients (54% women) referred because of clinically suspected PE between 2014 and 2018 were included. The age range was 16–100 years (median 72; interquartile range 18) and 25 female and 16 male patients underwent examinations twice, 1 female and 1 male patient underwent three examinations, and 1 female patient underwent four examinations (Table 1). The examinations were performed using five different CT scanners from three different manufacturers (Table 2). Ground truth for the measurements were generated through a comprehensive reading by the senior radiologist (TF) and the radiology resident (DT). For system development, 180 CTPA exams were randomly assigned to a training set and 520 were used for testing. Following quality scoring by the radiologist, 65% of training and 71% of test set cases, were of good or acceptable quality, whereas 35% and 29%, respectively, were classified as of inferior quality. No examinations were excluded.
A fully automatic system for detection, segmentation and measurement of mediastinal structures was developed (Figs. 1 and Supplemental Fig. 1). It was then trained using 180 CTPA examinations and tested on 520 CTPA examinations. The detection and segmentation outputs from both the training and the test datasets were independently evaluated by two radiologists (Table 3) with high initial inter-observer agreement (99.52–100%). For the very few cases where there was a disagreement, a reevaluation was made by both radiologists to reach a consensus evaluation. Evaluation of anatomical landmark detection in test cases showed that the tracheal bifurcation and the carina of the trachea were correctly located in 96% of the examinations while the pulmonary vein (PV)/proximal part of the PT was correctly located in 87%. Similarly, for cardiovascular structures, the DAo was detected in 90%, the AAo was detected in 86%, and the PT was detected in 88% of the test set examinations. Once compartments were correctly detected (Supplemental Fig. 16), a quantitative evaluation revealed successful segmentation of the trachea, the DAo, the AAo and the PT in 100% of the test cases. When analyzed by examination quality, the tracheal bifurcation and the carina of the trachea were correctly detected in 97%, 97%, and 93%, and the PV/proximal part of the PT was correctly detected in 90%, 86%, and 85% of good, acceptable and inferior quality CTPA exams, respectively. The DAo was correctly detected in 92%, 91%, and 89%, the AAo in 88%, 85%, and 85%, and the PT in 90%, 86%, and 86% of good, acceptable and inferior quality examinations, respectively (Fig. 2). External validation of the AAo segmentation was performed in 12 CTPAs from the SegTHOR dataset22, where the median Dice score between the CADe system and manual segmentation was 0.92 ± 0.02 SD and the median Boundary F1 (BF) contour matching score was 1.0 (Fig. 3).
Flowchart of the CADe algorithm. (A) A total of 700 2 mm axial CTPA image stacks were exported from the PACS server. (B) In the pre-processing, the linear scale value and the curvature of the CT image were calculated. (C) The segmentation chain consisted of four steps, starting with trachea detection followed by DAo, AAo, and PT detection. (D) Noise assessment and measurements of mediastinal vascular structures were reported by the system. Graphical output of the system is shown in Supplemental Fig. 15.
Automatic segmentation of trachea and vascular structures in CTPA examinations. Segmentation results of the CADe software (red) in representative CTPA examinations deemed by the radiologist as of good, acceptable or inferior quality are shown in red.
External validation of automatic AAo segmentation measurement. Boundary F1 and dice scores of the proposed CADe system for AAo segmentation in the SegTHOR dataset (n = 12 CTPA exams). Median (red line), interquartile range (boxes), and outliers ( +).
Next, we compared the performance of the CADe system versus the radiologist in image noise assessment, IV contrast level measurement in the PT, AAo diameter measurement and PT diameter measurement. The measurements of the CADe system and the radiologist were highly correlated (Pearson’s r = 0.87 for image noise, n = 470; 0.99 for IV contrast in the PT, n = 455; 0.92 for AAo, n = 447; and 0.68 for PT diameter, n = 455) (p < 0.001) (Fig. 4A1–D1). The limits of agreement between the CADe system and the radiologist had mean differences of − 0.25 HU for image noise, 0.28 HU for the IV contrast level measurement in the PT, 0.51 mm for the AAo diameter, and − 3.20 mm for the PT diameter (Fig. 4A2–D2). The Bland–Altman analysis revealed mean differences of − 1.08% for image noise, 0.02% for the IV contrast level measurement in the PT, 1.68% for the AAo diameter, and − 11.06% for the PT diameter. The CTPA image quality slightly affected the measurement performance of the mediastinal vessel structures. The AAo diameter and PT diameter measurements on good quality CTPA examinations showed a stronger correlation between CADe system and manual measurements than those in examinations with inferior image quality (Supplemental Figs. 17–20). The percentage of incorrect CADe measurements greater than ± 1.96 SD in successfully detected compartments was 4% for image noise and AAo diameter, 5% for IV contrast in PT, and 7% for PT diameter (Supplemental Table 2 and Supplemental Fig. 21). While 83% of the AAo diameter measurements were within 0.0 mm and 2.0 mm, only 41% of the PT diameter measurements were within 2.0 mm or less (Supplemental Table 3). When analyzed by examination quality, 85%, 86%, and 76% of the AAo diameter measurements, and 48%, 40%, and 33% of the PT diameter measurements were within the error range of 0.0–2.0 mm for good, acceptable and inferior quality CTPA exams, respectively. The mean deviation between the CADe system and the radiologist was for good, acceptable, and inferior image quality, respectively, 1.39 mm, 1.22 mm, and 1.77 mm for the AAo diameter measurements, and 3.44 mm, 4.32 mm, and 4.15 mm for the PT diameter measurements (Supplemental Fig. 22).
Comparison between the automated CADe system and the radiologist. Automatic and manual measurements from test set cases were compared by regression analysis (top row, dashed regression lines), Bland–Altman plots of differences in radiodensity and diameters (middle row, limits of agreement from − 1.96 to + 1.96 SD) and Bland–Altman plots of percentage differences (bottom row, limits of agreement from − 1.96 to + 1.96 SD). (A) Image noise (n = 470 CTPA exams). (B) Intravenous contrast agent in PT (n = 455 CTPA exams). (C) Ascending aorta diameter (n = 447 CTPA exams). (D) Pulmonary trunk diameter (n = 455 CTPA exams). The quality of each CTPA examination was assessed by the radiologist as good (black diamonds), acceptable (blue circles), or inferior (red squares).
External validation of the PT diameter measurement was performed in 35 CTPA exams from the FUMPE dataset23 where the CADe system successfully measured 31 exams (Pearson’s r = 0.83, p < 0.001) (Fig. 5A). For PT diameter, the limits of agreement between the CADe system and FUMPE radiologist annotation had mean differences of − 2.60 mm and -9.03% (Fig. 5B and C). Thus, the measurements of the automatic system were essentially on par with the radiologist in both our test dataset as well as in an external validation dataset.
External validation of automatic PT diameter measurement. Automatic PT diameter measurement from the FUMPE dataset cases (n = 31 CTPA exams) was compared to the manual radiologist annotation from the original study by (A) regression analysis (dashed regression lines), (B) Bland–Altman plots of differences in mm (limits of agreement from − 1.96 to + 1.96 SD) and (C) Bland–Altman plots of percentage differences (limits of agreement from − 1.96 to + 1.96 SD).
The CADe system was implemented in MATLAB R2019b, and the source code is available at https://github.com/aictpa/GreatVesselsAnalysis. The CADe system was tested on a workstation with Intel Core i9-9900X processors (3.5 GHz). Compartments were determined correctly in 178/180 and 442/520 examinations in the training and test sets respectively, and the computational performance on these examinations is shown in Table 4. The mean number of slices per examination was 146 ± 18 [SD] and the mean uploading time of CTPA examinations to the system was 1.1 ± 0.17 s. The mean run times for detection, segmentation, and measurement of all compartments was 5.4 ± 1 s in serial mode, and 1.2 ± 0.26 s in parallel mode.
The growing number of examinations performed per radiologist is rapidly becoming a challenge for healthcare24,25. Automated systems performing accurate detection, segmentation and measurements in CT imagery could be a solution but have so far not had sufficient performance for clinical implementation. We here developed and tested a deterministic algorithm that automatically detects, segments, and measures mediastinal structures in non- ECG-gated CTPA examinations.
We observed high concordance between the fully automated CADe system and the clinical state-of-the-art in noise assessment and mediastinal vascular measurements. Prior work using traditional image processing techniques to detect the AAo in a smaller set of 90 CTPA examinations had a success rate of 93% with mean absolute difference of AAo diameter measurement between algorithm and radiologist of 1.85 mm21. Here, the developed algorithm was 86% successful in detecting the AAo, but the mean absolute difference between algorithm and radiologist was only 0.51 mm. A deep learning based system tested on 288 CTPAs had mean differences between algorithm and radiologist of -0.94 mm for the AAo diameter, and − 0.86 mm for the PT diameter20. Another deep learning system to segment multiple cardiovascular structures including AAo, DAo, and PA obtained an overall median Dice score of 0.82 in a validation dataset of 42 examinations17. Here, the median Dice score for AAo segmentation was 0.92 in the external validation dataset, which represents a considerable improvement. However, the state-of-the-art methods were tested only in small datasets, whereas our algorithm was tested on a large set of unselected examinations reflecting routine radiology. Altogether, this study advances the fully deterministic detection, segmentation, and measurement of mediastinal structures in CTPA examinations, which is of particular importance for medical device software implementations intended for clinical use.
In clinical practice, a considerable proportion of the CTPA examinations will contain artifacts or other factors that reduce examination quality which complicates manual as well as automatic analysis. However, the quality of examinations has not been in focus in previous CTPA CADe work17,18,21. The high median age of patients under investigation for PE entails a higher prevalence of age-related anatomical deviations in the chest and mediastinum; it is therefore essential to train CADe systems on unselected examinations that reflect this clinical reality. Here, the system performed slightly better on CTPA examinations with good image quality than those with acceptable and inferior image qualities. Interestingly, system performance was only marginally worse for detection of anatomical landmark and cardiovascular structures in the examinations deemed of inferior quality as compared to those of acceptable quality. Examination quality had most impact on PT segmentation, mainly due to streak artifacts from the superior vena cava. High IV contrast within mediastinal structures may cause streak artifacts, which can alter the appearance of adjacent organs and make it difficult for the CADe system to detect and segment the organs. Examination quality also had an impact on the accuracy of diameter measurements. Differences in scanner models and CTPA protocols could potentially impact CADe performance if it was implemented at another hospital. Here, the CADe consistency is supported by the high similarity between results when the external SegTHOR and FUMPE datasets were used to validate different aspects of the system. Taken together, the developed system performs well in internal as well as external datasets independently of examination quality.
A limitation of the system is that the order of segmentation of anatomical structures is sequential. As the DAo was found to be relatively homogeneous and easily detected by a computer, DAo segmentation was performed before AAo and PT segmentation. The primary advantage of this approach is full automation without any input from the radiologist, but the main disadvantage is that failure at any step in the segmentation chain affects downstream detection and segmentation.
Unlike probabilistic approaches, such as deep convolutional neural networks (DCNN), the proposed deterministic model does not necessitate pixel-wise annotations for segmentation tasks. Therefore, the primary contribution of this approach is the elimination of the need for expert-level annotations during model development. Additionally, unlike the black box problem present in probabilistic approaches, all segmentation and measurement errors or unexpected outcomes can be easily tracked due to the inherent transparency of deterministic systems. In future work, several aspects need to be addressed. First, to fully leverage the proposed CADe system, the results should be integrated into the PACS server. One of the simplest means for achieving this is by generating Grey Scale Presentation State (GSPS) objects for segmentation and measurement outputs. This would enable the analysis of results generated by GSPS objects using any universal image viewer available on the PACS server. Second, there are additional structures that can provide prognostic information or serve as diagnostic indicators, such as volumetric analysis of the heart chambers, the right to left ventricle diameter ratio, or contrast reflux into the inferior vena cava26. Third, significant abnormalities may occur in structures at other anatomical levels. These anomalies in neighboring sections of the structure of interest can be identified by creating a 3D segmentation mask through the training of a deep learning model, such as the 3D U-Net model. To generate a 2D mask for training the deep learning model the segmentation output from the proposed CADe system can be utilized. Finally, our internal dataset with its corresponding annotations comprises solely CTPA volumes. Nonetheless, it is worth noting that the system was capable of processing contrast-enhanced CT images from the SegTHOR dataset. Therefore, as part of our future work, we intend to expand our testing to a larger set of contrast-enhanced CT images.
To the best of our knowledge, this system represents the first CTPA algorithm developed using a large number of cases with an adequate representation of common artifacts and challenges encountered in clinical radiology. The automatic CADe system detected and segmented anatomical landmarks, measured vascular structures of interest and determined relevant parameters of image quality. With its excellent computational performance, the system enables automatic assessment of large image volumes. This could also be used for research to clarify the diagnostic and prognostic impact of morphological and geometric properties of mediastinal structures. The algorithm can serve as a foundation for further CT chest AI development and already in its current form provide radiologists with valuable automated measurements.
The dataset consists of 700 retrospective non-electrocardiogram (ECG)-gated CTPA examinations performed at a single institution (Nyköping Hospital, Nyköping, Sweden) between 2014 and 2018; 383 CTPA examinations from 353 women (age range 16–97; median age 73; interquartile range 20) and 317 from 299 men (age range 19–100; median age 71; interquartile range 15). The CTPAs were performed in four different time periods to include examinations from several different CT scanners. Consecutive CTPAs ensured an image material with adequate representation of gender, age, common artifacts and medical conditions. The examinations were randomly assigned to a training set (n = 180) used for algorithm development, and a test set (n = 520) for unbiased evaluation of the final algorithm. Collection and analysis of CTPA examinations was approved by the Swedish Ethical Review Authority (EPN Uppsala Dnr 2015/023 and 2015/023/1). All personal identifiers in Digital Imaging and Communications in Medicine (DICOM) headers were removed from the dataset (Dicom2usb). We validated the proposed computer-aided detection (CADe) system on two publicly available datasets: 12 CTPAs from Segmentation of Thoracic Organs at Risk (SegTHOR22) (AAo segmentation), and 35 CTPA exams from the Ferdowsi University of Mashhad’s PE (FUMPE23) dataset (PT diameter measurement).
The non-ECG-gated CTPA examinations were conducted with 5 different multidetector-row CT scanners (Brilliance 64, Ingenuity Core and Ingenuity CT, Philips Medical Systems; LightSpeed VCT, General Electric (GE) Healthcare Systems; Somatom Definition Flash, Siemens Healthcare) after IV injection of contrast (Omnipaque 350 mg I/ml, GE Healthcare Systems) and saline. The CT image acquisition technique varied by manufacturer with most frequent slice thickness of 0.625 mm (0.625 mm–2.0 mm), pixel spacing of 0.7 mm (0.59 mm–0.98 mm), and voltage of 100 kV (80 kV–120 kV). A secondary axial reformat with 2.0 mm slice thickness was performed on all examinations.
The CTPA examinations were exported from the Picture Archiving and Communication System (PACS, Sectra AB) in DICOM format. The examinations were reviewed and annotated using the RadiAnt DICOM Viewer (Medixant) by a senior radiologist (TF) with 15 years of experience. Of the CTPAs, 150 were first reviewed and annotated by a radiology resident (DT) with 5 years of experience in general diagnostic radiology and then double read by TF. In the 2 mm axial image stack, the image which optimally presented the PT was identified and the diameter of the PT and AAo, IV contrast concentration in PT (mean value of HU in 2 cm2 circular ROI), and image noise (SD of HU in a 1 cm2 circular ROI in the DAo) were measured by the radiologist and used as ground truth (Fig. 6). For each CTPA examination the radiologist also scored five image quality parameters affecting the evaluation for PE: motion artifacts, streak artifacts, IV contrast concentration in PT, parenchymal disease, and image noise (Supplemental Methods: Image score calculations and Supplemental Table 1). A total score was calculated, and the result determined the overall CTPA examination quality as good (score 0–3), acceptable (4–7) or inferior (≥ 8) (Fig. 7).
Ground truth measurements by the radiologist. The measurements were performed in the axial 2 mm image which optimally presented the PT. (A) The IV contrast concentration in the PT was recorded as the mean HU value in a 2 cm2 circular region of interest (arrow). Image noise was determined as the SD of HU in a 1 cm2 circular region of interest in the DAo (arrowhead). (B). The diameters of the PT (arrow) and AAo (arrowhead) were measured.
Representative examples of CTPA examinations of different image quality. The examinations were classified by the radiologist as being of good, acceptable or inferior quality. The levels of the carina of the trachea, the PT and the PV are shown with the same window setting (width = 600 HU, level = 100 HU).
To automatically detect, segment and measure mediastinal structures (AAo, DAo, PT), we combined image processing and image analysis techniques to detect and segment compartments. The developed CADe system does not require any user interaction and consists of two main steps, pre-processing and a segmentation chain (Fig. 1 and Supplemental Fig. 1).
For each axial CT image, every voxel was converted to HU and the direction of scanning was determined based on information in the DICOM header. As patient orientation may vary along the cranial to caudal direction because of scoliosis, movement or position during the examination, or other reasons, the CT exam was aligned with respect to the x-axis of the axial plane in the cranial as well as caudal part of the examination (Fig. 1B1 and B2, Supplemental methods: Description of the CADe system 1–4, and Supplemental Figs. 2–6).
First, the system located the three-dimensional positions of the carina of trachea and an apical level of the pulmonary valve (PV), to find seed points of vascular structures in the mediastinum. Second, the seed points for segmentation were automatically placed using a heuristic approach. Third, the structures were segmented by applying image processing techniques for image enhancement, edge detection, gray scale segmentation, and 2D region growing. Finally, measurements were done on the segmented regions (Supplemental Figs. 7–14).
Details are found in Supplemental methods: Description of the CADe system 5–8. The CADe system extracts two adjoined volumes of interest from the superior part of the thoracic cavity, wherein trachea candidates are generated and assessed separately (Fig. 1C4 and C5). In these two volumes, three air-filled structures (the trachea and the left and right lung) are identified by thresholding, flood-fill operations and connected component analyses. Trachea candidates are then joined across the two volumes and connected component analyses are applied. The longest of the candidates also having diameter and volume within an empirically determined range is selected as the trachea. When the trachea has been detected, it is tracked cranially to caudally slice by slice to the bifurcation point where the left and right main bronchi can be found as two distinct segments (Fig. 1C6). We then designate the CT slice, where the distance between the left and right main bronchus is > 0.75 cm, as the level of the carina of the trachea. Thus, the trachea and the carina of the trachea can be located automatically in 3D.
Details are found in Supplemental methods: Description of the CADe system 9–16. We observed that the DAo can be easily detected around the level of carina trachea since the DAo is always located posterior to the left main bronchus and its appearance (IV contrast concentration and circularity) is homogeneous around the level of carina trachea. Two sequential sets of artificial ray search spaces were made. One or more rays passing a sufficient number of connected pixels to correspond to the contrast filled aorta indicate that the DAo has been located (Fig. 1C9). Once the DAo was detected, a segmentation chain was applied to detect the aortic arch. The area of a 1 cm2 circle at the mass center of the DAo was used to calculate the average HU density of the CTPA examination as a metric of contrast filling (Fig. 1C12). The AAo can be detected by tracking and comparing the segmented DAo regions slice by slice in the caudal to cranial direction to find features characteristic of the AAo. Morphological changes indicate the level of the aortic arch (Fig. 1C13). By tracking the anterior part of the aortic arch in the cranial to caudal direction slice by slice the first identified circular object was designated as AAo (Fig. 1C14). Next, we segmented the AAo between the levels of the aortic arch and the carina of trachea (Fig. 1C15) and calculated the diameter of AAo as the mean of AAo diameters in these planes. The PT is adjacent to the AAo and remains on its left. We therefore created a rectangular search space adjacent to the left lateral side of AAo (Fig. 1C18) and applied our segmentation pipeline to this search space to segment the PT. Next, the PT was tracked in the cranial-to-caudal direction to reach the PV/proximal part of the PT. The circularity of the segmented region was used to determine whether the level of the PV had been reached (Fig. 1C20). The tracked PT diameter was calculated by Hough transform (Fig. 1C20) as the mean of PT diameters in these planes. Taken together, the DAo, AAo and PT were automatically detected in 3D and their average diameters and contrast levels obtained.
Statistical analysis was performed using Microsoft Office Excel (Microsoft Corporation, Office Professional Plus 2016). Bland–Altman and scatter analyses were used to compare agreement and relationship, respectively, between automated and manual measurements. A p-value of < 0.05 was defined as statistically significant and Pearson’s correlation coefficient was used to evaluate the agreement between automated measurement and manual measurement. The accuracy of the CADe system was assessed by Boundary F1 score and the Dice coefficient score using Matlab (MathWorks, Inc., R2019b).
The datasets generated during and/or analysed during the current study are available from the corresponding author on reasonable request.
The developed model code can be found at the following link. https://github.com/aictpa/GreatVesselsAnalysis.
Rubin, G. D., Leipsic, J., Schoepf, U. J., Fleischmann, D. & Napel, S. CT angiography after 20 years: A transformation in cardiovascular disease characterization continues to advance. Radiology 271, 633–652 (2014).
Konstantinides, S. V. et al. 2014 ESC guidelines on the diagnosis and management of acute pulmonary embolism: The task force for the diagnosis and management of acute pulmonary embolism of the european society of cardiology (ESC) endorsed by the european respiratory society (ERS). Eur. Heart J. 35, 3033–3080 (2014).
Article CAS PubMed Google Scholar
Hu, X. et al. Use of pulmonary CT angiography with low tube voltage and low-iodine-concentration contrast agent to diagnose pulmonary embolism. Sci. Rep. 7, 1–8 (2017).
Sista, A. K., Kuo, W. T., Schiebler, M. & Madoff, D. C. Stratification, imaging, and management of acute massive and submassive pulmonary embolism. Radiology 284, 5–24 (2017).
Office of the Surgeon General (US) & National Heart, Lung, and Blood Institute (US). The Surgeon General’s Call to Action to Prevent Deep Vein Thrombosis and Pulmonary Embolism (Office of the Surgeon General (US)) (2008).
Abadi, E., Sanders, J. & Samei, E. Patient-specific quantification of image quality: An automated technique for measuring the distribution of organ Hounsfield units in clinical chest CT images. Med. Phys. 44, 4736–4746 (2017).
Electronic Code of Federal Regulations (eCFR). https://www.ecfr.gov/current/title-21/chapter-I/subchapter-J/part-1020#1020.33.
Bae, K. T. et al. CT depiction of pulmonary emboli: Display window settings. Radiology 236, 677–684 (2005).
Jinlin, Wu. et al. Ascending aortic length and risk of aortic adverse events. J. Am. Coll. Cardiol. 74, 1883–1894 (2019).
Lavall, D., Schäfers, H.-J., Böhm, M. & Laufs, U. Aneurysms of the ascending aorta.German.Doctor's noteInt.109, 227–233 (2012).
PubMed PubMed Central Google Scholar
Aluja Jaramillo, F. et al. Approach to pulmonary hypertension: From CT to clinical diagnosis. Radiographics 38, 357–373 (2018).
Mahammedi, A., Oshmyansky, A., Hassoun, P. M., Thiemann, D. R. & Siegelman, S. S. Pulmonary artery measurements in pulmonary hypertension: The role of computed tomography. J. Thorac. Imaging 28, 96–103 (2013).
Aviram, G. et al. Pulmonary hypertension: A nomogram based on CT pulmonary angiographic data for prediction in patients without pulmonary embolism. Radiology 277, 236–246 (2015).
Meinel, F. G. et al. Predictive value of computed tomography in acute pulmonary embolism: Systematic review and meta-analysis. Am. J. Med. 128, 747–759 (2015).
Kang, D. K. et al. CT signs of right ventricular dysfunction: Prognostic role in acute pulmonary embolism. J. Am. Coll. Cardiol. Imaging 4, 841–849 (2011).
Saur, SC, Kühhnel, C., Boskamp, T., Székely, G. & Cattin, P. Automatic ascending aorta detection in CTA datasets.In Image Processing for Medicine 2008 (eds Tolxdorff, T. et al.) 323–327 (Springer, 2008).
Baskaran, L. et al. Automatic segmentation of multiple cardiovascular structures from cardiac computed tomography angiography images using deep learning. PloS One 15, e0232573 (2020).
Article CAS PubMed PubMed Central Google Scholar
Ecabert, O. et al. Segmentation of the heart and great vessels in CT images using a model-based adaptation framework. Med. Image Anal. 15, 863–876 (2011).
Wang, S., Fu, L., Yue, Y., Kang, Y. & Liu, J. Fast and automatic segmentation of ascending aorta in MSCT volume data. In 2009 2nd International Congress on Image and Signal Processing (eds Wang, S. et al.) 1–5 (IEEE, 2009).
Chettrit, D., Amitai, O. B., Tamir, I., Bar, A. & Elnekave, E. PHT-bot: A deep learning based system for automatic risk stratification of COPD patients based upon signs of pulmonary hypertension. In Medical Imaging 2019: Computer-Aided Diagnosis Vol. 10950 (eds Chettrit, D. et al.) (International Society for Optics and Photonics, 2019).
Pfeffer, Y., Mayer, A., Zholkover, A. & Konen, E. A system for automatic aorta sections measurements on chest CT. In Medical Imaging 2016: Computer-Aided Diagnosis Vol. 9785 (eds Pfeffer, Y. et al.) 978505 (International Society for Optics and Photonics, 2016).
Lambert, Z., Petitjean, C., Dubray, B. & Kuan, S. SegTHOR: Segmentation of thoracic organs at risk in CT images. In 2020 Tenth International Conference on Image Processing Theory, Tools and Applications (IPTA), 1–6 (IEEE, 2020).
Masoudi, M. et al. A new dataset of computed-tomography angiography images for computer-aided detection of pulmonary embolism. Sci. Data 5, 180180 (2018).
Article PubMed PubMed Central Google Scholar
Royal College of Radiologists. Clinical radiology UK workforce census 2016 report, 1–57 (2016).
OECD iLibrary | Health at a Glance 2017: OECD indicators. https://www.oecd-ilibrary.org/social-issues-migration-health/health-at-a-glance-2017_health_glance-2017-en.
Konstantinides, S. V. et al. 2019 ESC guidelines for the diagnosis and management of acute pulmonary embolism developed in collaboration with the European respiratory society (ERS). Eur. Heart J. 41, 543–603 (2020).
The project was supported by a grant from Analytic Imaging Diagnostic Arena (AIDA) to TS. TF and DT were financially supported by clinical fellowships from AIDA. TF was financially supported by the Centre for Clinical Research Sörmland, Uppsala University.
Open access funding provided by Uppsala University. The project was supported by a grant from Analytic Imaging Diagnostic Arena (AIDA), https://medtech4health.se/aida-en/, to TS. TF and DT were supported by clinical fellowships from AIDA. TF was supported by the Centre for Clinical Research Sörmland, Uppsala University, Eskilstuna, Sweden, VINNOVA.
Department of Immunology, Genetics and Pathology, Uppsala University, 751 85, Uppsala, Sweden
Ali T. Kahraman & Tobias Sjöblom
Department of Radiology, Nyköping Hospital, 611 39, Nyköping, Sweden
Department of Radiology, Uppsala University Hospital, 751 85, Uppsala, Sweden
Department of Surgical Sciences, Uppsala University, 751 85, Uppsala, Sweden
Centre for Image Analysis, Department of Information Technology, Uppsala University, 751 05, Uppsala, Sweden
You can also search for this author in PubMed Google Scholar
You can also search for this author in PubMed Google Scholar
You can also search for this author in PubMed Google Scholar
You can also search for this author in PubMed Google Scholar
You can also search for this author in PubMed Google Scholar
A.T.K., T.F. and T.S. designed the study. A.T.K. developed the theory and performed implementations and the computations. A.T.K. wrote the manuscript with support from T.F., N.S. and T.S. N.S. analyzed the developed model. T.F. collected data. T.F. and D.T. prepared the ground truth data and validated the results. All authors discussed the results, provided critical feedback and contributed to the final manuscript. T.S. supervised the project.
Correspondence to Tobias Sjöblom.
The authors declare no competing interests.
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
Kahraman, A.T., Fröding, T., Toumpanakis, D. et al. Automated detection, segmentation and measurement of major vessels and the trachea in CT pulmonary angiography. Sci Rep 13, 18407 (2023). https://doi.org/10.1038/s41598-023-45509-1
DOI: https://doi.org/10.1038/s41598-023-45509-1
Anyone you share the following link with will be able to read this content:
Sorry, a shareable link is not currently available for this article.
Provided by the Springer Nature SharedIt content-sharing initiative
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.
Scientific Reports (Sci Rep) ISSN 2045-2322 (online)
High Accuracy Measurement Sign up for the Nature Briefing newsletter — what matters in science, free to your inbox daily.